Dose enhancement from backscattered radiation at tissue–metal interfaces irradiated with high energy electrons

Abstract

Irradiation of a tissue–metal interface with 13 MeV to 20 MeV electrons results in an increased dose to the tissue on the entrance side of the metal. Ionization measurements were made with a thin-window parallel-plate chamber to determine the magnitude of the dose enhancement as a function of incident electron energy, thickness and atomic number of metals introduced into the electron beam. The presence of a metal resulted in a dose ranging from 6% to approximately 50% greater than that measured with no metal in the beam. Most of this increase in dose may be eliminated by the addition of 1–2 g/cm2 of low Z material between tissue and metal.

Volume 53, Issue 629May 1980